volume 16 issue 3 publication number e202200288

Feasibility test of a sapphire cryoprobe with optical monitoring of tissue freezing

Arsen K. Zotov 1
Larisa P. Safonova 2
Polina V. Aleksandrova 3
Igor V. Reshetov 4
Kirill I. Zaytsev 3
Valery V. Tuchin 5, 6, 7
Vladimir N. Kurlov 1
Publication typeJournal Article
Publication date2022-12-20
scimago Q2
wos Q3
SJR0.531
CiteScore4.7
Impact factor2.3
ISSN1864063X, 18640648
General Chemistry
General Biochemistry, Genetics and Molecular Biology
General Physics and Astronomy
General Materials Science
General Engineering
Abstract

This article describes a sapphire cryoprobe as a promising solution to the significant problem of modern cryosurgery that is the monitoring of tissue freezing. This probe consists of a sapphire rod manufactured by the edge‐defined film‐fed growth technique from Al2O3 melt and optical fibers accommodated inside the rod and connected to the source and the detector. The probe's design enables detection of spatially resolved diffuse reflected intensities of tissue optical response, which are used for the estimation of tissue freezing depth. The current type of the 12.5‐mm diameter sapphire probe cooled down by the liquid nitrogen assumes a superficial cryoablation. The experimental test made by using a gelatin‐intralipid tissue phantom shows the feasibility of such concept, revealing the capabilities of monitoring the freezing depth up to 10 mm by the particular instrumentation realization of the probe. This justifies a potential of sapphire‐based instruments aided by optical diagnosis in modern cryosurgery.

Found 
Found 

Top-30

Journals

1
Crystals
1 publication, 33.33%
Sensors
1 publication, 33.33%
Journal of Biophotonics
1 publication, 33.33%
1

Publishers

1
2
MDPI
2 publications, 66.67%
Wiley
1 publication, 33.33%
1
2
  • We do not take into account publications without a DOI.
  • Statistics recalculated weekly.

Are you a researcher?

Create a profile to get free access to personal recommendations for colleagues and new articles.
Metrics
3
Share
Cite this
GOST |
Cite this
GOST Copy
Dolganova I. N. et al. Feasibility test of a sapphire cryoprobe with optical monitoring of tissue freezing // Journal of Biophotonics. 2022. Vol. 16. No. 3. e202200288
GOST all authors (up to 50) Copy
Dolganova I. N., Zotov A. K., Safonova L. P., Aleksandrova P. V., Reshetov I. V., Zaytsev K. I., Tuchin V. V., Kurlov V. N. Feasibility test of a sapphire cryoprobe with optical monitoring of tissue freezing // Journal of Biophotonics. 2022. Vol. 16. No. 3. e202200288
RIS |
Cite this
RIS Copy
TY - JOUR
DO - 10.1002/jbio.202200288
UR - https://onlinelibrary.wiley.com/doi/10.1002/jbio.202200288
TI - Feasibility test of a sapphire cryoprobe with optical monitoring of tissue freezing
T2 - Journal of Biophotonics
AU - Dolganova, Irina N.
AU - Zotov, Arsen K.
AU - Safonova, Larisa P.
AU - Aleksandrova, Polina V.
AU - Reshetov, Igor V.
AU - Zaytsev, Kirill I.
AU - Tuchin, Valery V.
AU - Kurlov, Vladimir N.
PY - 2022
DA - 2022/12/20
PB - Wiley
IS - 3
VL - 16
PMID - 36510652
SN - 1864-063X
SN - 1864-0648
ER -
BibTex
Cite this
BibTex (up to 50 authors) Copy
@article{2022_Dolganova,
author = {Irina N. Dolganova and Arsen K. Zotov and Larisa P. Safonova and Polina V. Aleksandrova and Igor V. Reshetov and Kirill I. Zaytsev and Valery V. Tuchin and Vladimir N. Kurlov},
title = {Feasibility test of a sapphire cryoprobe with optical monitoring of tissue freezing},
journal = {Journal of Biophotonics},
year = {2022},
volume = {16},
publisher = {Wiley},
month = {dec},
url = {https://onlinelibrary.wiley.com/doi/10.1002/jbio.202200288},
number = {3},
pages = {e202200288},
doi = {10.1002/jbio.202200288}
}