Journal of Prosthetic Dentistry, volume 53, issue 6, pages 800-804
Accuracy of an earpiece face-bow
Joyce F Palik
1
,
Donald R. Nelson
1
,
James T. White
1
Publication type: Journal Article
Publication date: 1985-06-01
Journal:
Journal of Prosthetic Dentistry
scimago Q1
SJR: 1.177
CiteScore: 7.0
Impact factor: 4.3
ISSN: 00223913, 10976841
PubMed ID:
3859649
Oral Surgery
Abstract
Abstract
The validity of the Hanau ear-bow to transfer an arbitrary hinge axis to a Hanau articulator was clinically compared with a Hanau kinematic face-bow. The study was conducted with 18 randomly selected patients. This investigation demonstrated a significant statistical difference between the arbitrary axis located with an ear-bow and the terminal hinge axis. This discrepancy was significant in the anteroposterior direction but not in the superior-inferior direction. Only 50% of the arbitrary hinge axes were within a 5 mm radius of the terminal hinge axis, while 89% were within a 6 mm radius. Furthermore, the ear-bow method was not repeatable statistically. Additional study is needed to determine the practical value of the arbitrary face-bow and to pursue modifications to improve its accuracy.Found