Misaligned mechanical ultrasound transducer contributes to carotid puncture
Publication type: Journal Article
Publication date: 2007-12-10
SJR: —
CiteScore: —
Impact factor: —
ISSN:
Anesthesiology and Pain Medicine
Abstract
This report describes a complication from the use of a mechanical ultrasound transducer and needle guide to perform internal jugular cannulation. Misalignment of the probe led to an inability to visualize the needle and accidental carotid puncture. Placement of the central line was successful with landmark technique, and there were no sequelae. The transducer design explains how such misalignment may occur if the crystal oscillation is not symmetric. The manufacturer recommends periodic testing for probe alignment, and awareness of this potential problem may help other users avoid patient injury with this type of ultrasound transducer.
Found
Nothing found, try to update filter.
Found
Nothing found, try to update filter.
Top-30
Journals
|
1
|
|
|
Canadian Journal of Anaesthesia
1 publication, 50%
|
|
|
Proceedings of SPIE - The International Society for Optical Engineering
1 publication, 50%
|
|
|
1
|
Publishers
|
1
|
|
|
Springer Nature
1 publication, 50%
|
|
|
SPIE-Intl Soc Optical Eng
1 publication, 50%
|
|
|
1
|
- We do not take into account publications without a DOI.
- Statistics recalculated weekly.
Are you a researcher?
Create a profile to get free access to personal recommendations for colleagues and new articles.
Metrics
2
Total citations:
2
Citations from 2024:
0
Cite this
GOST |
RIS |
BibTex |
MLA
Cite this
GOST
Copy
Weller R. O. Misaligned mechanical ultrasound transducer contributes to carotid puncture // Seminars in Anesthesia, Perioperative Medicine and Pain. 2007. Vol. 26. No. 4. pp. 193-196.
GOST all authors (up to 50)
Copy
Weller R. O. Misaligned mechanical ultrasound transducer contributes to carotid puncture // Seminars in Anesthesia, Perioperative Medicine and Pain. 2007. Vol. 26. No. 4. pp. 193-196.
Cite this
RIS
Copy
TY - JOUR
DO - 10.1053/j.sane.2007.08.001
UR - https://doi.org/10.1053/j.sane.2007.08.001
TI - Misaligned mechanical ultrasound transducer contributes to carotid puncture
T2 - Seminars in Anesthesia, Perioperative Medicine and Pain
AU - Weller, R. O.
PY - 2007
DA - 2007/12/10
PB - Elsevier
SP - 193-196
IS - 4
VL - 26
SN - 0277-0326
ER -
Cite this
BibTex (up to 50 authors)
Copy
@article{2007_Weller,
author = {R. O. Weller},
title = {Misaligned mechanical ultrasound transducer contributes to carotid puncture},
journal = {Seminars in Anesthesia, Perioperative Medicine and Pain},
year = {2007},
volume = {26},
publisher = {Elsevier},
month = {dec},
url = {https://doi.org/10.1053/j.sane.2007.08.001},
number = {4},
pages = {193--196},
doi = {10.1053/j.sane.2007.08.001}
}
Cite this
MLA
Copy
Weller, R. O.. “Misaligned mechanical ultrasound transducer contributes to carotid puncture.” Seminars in Anesthesia, Perioperative Medicine and Pain, vol. 26, no. 4, Dec. 2007, pp. 193-196. https://doi.org/10.1053/j.sane.2007.08.001.