Journal of wound care, volume 28, issue 7, pages 438-443

Efficacy of a bacterial fluorescence imaging device in an outpatient wound care clinic: a pilot study

Publication typeJournal Article
Publication date2019-07-02
scimago Q2
SJR0.404
CiteScore2.9
Impact factor1.5
ISSN09690700, 20522916
Fundamentals and skills
Nursing (miscellaneous)
Abstract
Objective:

Subsurface bacterial burden can be missed during standard wound examination protocols. The real-time bacterial fluorescence imaging device, MolecuLight i:X, visualises the presence of potentially harmful levels of bacteria through endogenous autofluorescence, without the need for contrast agents or contact with the patient. The intended use of the imaging device is to assist with the management of patients with wounds by enabling real-time visualisation of potentially harmful bacteria. The aim of this study was to establish the accuracy of the wound imaging device at detecting pathogenic bacteria in wounds.

Methods:

A single-centre, prospective observational study was conducted in Cork University Hospital in an outpatient plastic surgery wound care clinic. Patients had their wounds photographed under white and autofluorescent light with the imaging device. Auto-fluorescent images were compared with the microbiological swab results.

Results:

A total of 33 patients and 43 swabs were included, of which 95.3% (n=41) were positive for bacteria growth. Staphylococcus aureus was the most common bacterial species identified. The imaging device had a sensitivity of 100% and specificity of 78% at identifying pathological bacteria presence in wounds on fluorescent light imaging. The positive predictive value (PPV) was 95.4%. The negative predictive value (NPV) was 100%. It demonstrated a sensitivity and specificity of 100% at detecting the presence of Pseudomonas spp.

Conclusion:

The imaging device used could be a safe, effective, accurate and easy-to-use autofluorescent device to improve the assessment of wounds in the outpatient clinic setting. In conjunction with best clinical practice, the device can be used to guide clinicians use of antibiotics and specialised dressings.

Top-30

Journals

1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8

Publishers

2
4
6
8
10
2
4
6
8
10
  • We do not take into account publications without a DOI.
  • Statistics recalculated only for publications connected to researchers, organizations and labs registered on the platform.
  • Statistics recalculated weekly.

Are you a researcher?

Create a profile to get free access to personal recommendations for colleagues and new articles.
Share
Cite this
GOST | RIS | BibTex | MLA
Found error?