Applied Sciences (Switzerland), volume 11, issue 15, pages 7166

The Influence of Hard- and Software Improvement of Intraoral Scanners on the Implant Transfer Accuracy from 2012 to 2021: An In Vitro Study

Alexander Schmidt 1
Maximiliane Amelie Schlenz 1
Haoyu Liu 1
Holger Sebastian Kämpe 2
Bernd Wöstmann 1
Publication typeJournal Article
Publication date2021-08-03
scimago Q2
SJR0.508
CiteScore5.3
Impact factor2.5
ISSN20763417
Computer Science Applications
Process Chemistry and Technology
General Materials Science
Instrumentation
General Engineering
Fluid Flow and Transfer Processes
Abstract

This study aimed to investigate the transfer accuracy (trueness and precision) of three different intraoral scanning families using different hardware and software versions over the last decade from 2012 to 2021, compared to a conventional impression. Therefore, an implant master model with a reference cube was digitized and served as a reference dataset. Digital impressions of all three scanning families (True definition, TRIOS, CEREC) were recorded (n = 10 per group), and conventional implant impressions were taken (n = 10). The conventional models were digitized, and all models (conventional and digital) were measured. Therefore, it was possible to obtain the deviations between the master model and the scans or conventional models in terms of absolute three-dimensional (3D) deviations, deviations in rotation, and angulation. The results for deviations between the older and newer scanning systems were analyzed using pairwise comparisons (p < 0.05; SPSS 26). The absolute 3D deviations increased with increasing scan path length, particularly for the older hardware and software versions (old vs. new (MW ± SD) True Definition: 355 ± 62 µm vs. 483 ± 110 µm; TRIOS: 574 ± 274 µm vs. 258 ± 100 µm; and CEREC: 1356 ± 1023 µm vs. 110 ± 49 µm). This was also true for deviations in rotation and angulation. The conventional impression showed an advantage only regarding the absolute 3D deviation compared to the older systems. Based on the data of the present study, the accuracy of intraoral scanners is decisively related to hardware and software; though, newer systems or software do not necessarily warrant improvement. Nevertheless, to achieve high transfer accuracy, regular updating of digital systems is recommended. The challenge of increasing errors with increasing scan paths is overcome in the most recent systems. The combination of two different scanning principles in a single device seems to be beneficial.

Found 
Found 

Top-30

Journals

1
2
1
2

Publishers

1
2
3
1
2
3
  • We do not take into account publications without a DOI.
  • Statistics recalculated only for publications connected to researchers, organizations and labs registered on the platform.
  • Statistics recalculated weekly.

Are you a researcher?

Create a profile to get free access to personal recommendations for colleagues and new articles.
Share
Cite this
GOST | RIS | BibTex | MLA
Found error?