A New Multi-Axial Functional Stress Analysis Assessing the Longevity of a Ti-6Al-4V Dental Implant Abutment Screw

Ghada H Naguib 1, 2
Ahmed O. Abougazia 3
Lulwa E. Al-Turki 4
Hisham A Mously 4
Abou Bakr Hossam Hashem 5
Abdulghani I. Mira 6
Osama A. Qutub 4
Abdulelah M. Binmahfooz 4
Afaf A. Almabadi 4
Mohamed T Hamed 4, 7
Publication typeJournal Article
Publication date2024-11-12
scimago Q2
wos Q1
SJR0.647
CiteScore4.2
Impact factor3.9
ISSN23137673
Abstract

This study investigates the impact of tightening torque (preload) and the friction coefficient on stress generation and fatigue resistance of a Ti-6Al-4V abutment screw with an internal hexagonal connection under dynamic multi-axial masticatory loads in high-cycle fatigue (HCF) conditions. A three-dimensional model of the implant–abutment assembly was simulated using ANSYS Workbench 16.2 computer aided engineering software with chewing forces ranging from 300 N to 1000 N, evaluated over 1.35 × 107 cycles, simulating 15 years of service. Results indicate that the healthy range of normal to maximal mastication forces (300–550 N) preserved the screw’s structural integrity, while higher loads (≥800 N) exceeded the Ti-6Al-4V alloy’s yield strength, indicating a risk of plastic deformation under extreme conditions. Stress peaked near the end of the occluding phase (206.5 ms), marking a critical temporal point for fatigue accumulation. Optimizing the friction coefficient (0.5 µ) and preload management improved stress distribution, minimized fatigue damage, and ensured joint stability. Masticatory forces up to 550 N were well within the abutment screw’s capacity to sustain extended service life and maintain its elastic behavior.

Found 
Found 

Top-30

Journals

1
Frontiers in Bioengineering and Biotechnology
1 publication, 50%
BMC Oral Health
1 publication, 50%
1

Publishers

1
Frontiers Media S.A.
1 publication, 50%
Springer Nature
1 publication, 50%
1
  • We do not take into account publications without a DOI.
  • Statistics recalculated weekly.

Are you a researcher?

Create a profile to get free access to personal recommendations for colleagues and new articles.
Metrics
2
Share
Cite this
GOST |
Cite this
GOST Copy
Naguib G. H. et al. A New Multi-Axial Functional Stress Analysis Assessing the Longevity of a Ti-6Al-4V Dental Implant Abutment Screw // Biomimetics. 2024. Vol. 9. No. 11. p. 689.
GOST all authors (up to 50) Copy
Naguib G. H., Abougazia A. O., Al-Turki L. E., Mously H. A., Hashem A. B. H., Mira A. I., Qutub O. A., Binmahfooz A. M., Almabadi A. A., Hamed M. T. A New Multi-Axial Functional Stress Analysis Assessing the Longevity of a Ti-6Al-4V Dental Implant Abutment Screw // Biomimetics. 2024. Vol. 9. No. 11. p. 689.
RIS |
Cite this
RIS Copy
TY - JOUR
DO - 10.3390/biomimetics9110689
UR - https://www.mdpi.com/2313-7673/9/11/689
TI - A New Multi-Axial Functional Stress Analysis Assessing the Longevity of a Ti-6Al-4V Dental Implant Abutment Screw
T2 - Biomimetics
AU - Naguib, Ghada H
AU - Abougazia, Ahmed O.
AU - Al-Turki, Lulwa E.
AU - Mously, Hisham A
AU - Hashem, Abou Bakr Hossam
AU - Mira, Abdulghani I.
AU - Qutub, Osama A.
AU - Binmahfooz, Abdulelah M.
AU - Almabadi, Afaf A.
AU - Hamed, Mohamed T
PY - 2024
DA - 2024/11/12
PB - MDPI
SP - 689
IS - 11
VL - 9
PMID - 39590261
SN - 2313-7673
ER -
BibTex |
Cite this
BibTex (up to 50 authors) Copy
@article{2024_Naguib,
author = {Ghada H Naguib and Ahmed O. Abougazia and Lulwa E. Al-Turki and Hisham A Mously and Abou Bakr Hossam Hashem and Abdulghani I. Mira and Osama A. Qutub and Abdulelah M. Binmahfooz and Afaf A. Almabadi and Mohamed T Hamed},
title = {A New Multi-Axial Functional Stress Analysis Assessing the Longevity of a Ti-6Al-4V Dental Implant Abutment Screw},
journal = {Biomimetics},
year = {2024},
volume = {9},
publisher = {MDPI},
month = {nov},
url = {https://www.mdpi.com/2313-7673/9/11/689},
number = {11},
pages = {689},
doi = {10.3390/biomimetics9110689}
}
MLA
Cite this
MLA Copy
Naguib, Ghada H., et al. “A New Multi-Axial Functional Stress Analysis Assessing the Longevity of a Ti-6Al-4V Dental Implant Abutment Screw.” Biomimetics, vol. 9, no. 11, Nov. 2024, p. 689. https://www.mdpi.com/2313-7673/9/11/689.