Light Output Function and Pulse-Shape Discrimination Capability of p-Terphenyl Organic Scintillator in Wide Neutron Energy Range of 1.1 to 19 MeV

A Jančář 1
Zdeněk Kopecký 1
Jiří Čulen 1
Zdeněk Matěj 2
Michal Košťál 3
1
 
VF NUCLEAR, Svitavská 588, 679 21 Černá Hora, Czech Republic
3
 
Research Centre Řež Ltd., 130, 250 68 Prague, Czech Republic
Publication typeJournal Article
Publication date2024-10-22
Journal: Instruments
scimago Q2
SJR0.658
CiteScore2.6
Impact factor
ISSN2410390X
Abstract

In this work, we studied the light-output properties, efficiency function, as well as the pulse-shape discrimination (PSD) capability of p-Terphenyl scintillator. The selected solid cylindrical scintillation detector has a thickness of 45 mm and a diameter of 45 mm. Recently presented studies of light-output functions have only been measured for low-neutron energies. Our motivation has been to determine the light output function for p-Terphenyl scintillator more accurately over a wider neutron energy range. The measurements have been carried out with mono-energetic neutron beams in the wide energy range from 1.1 to 19 MeV. The neutron–gamma spectrometric system which we developed has been used for the measurement. The input analog signal from the detector was digitized with a fast 12-bits analog to digital converter with a sampling frequency of 1 GHz. Measured data from the detector are processed into the gamma and neutron spectra. The accurate light output function for the p-Therphenyl scintillator has been calculated. The pulse-shape discrimination capability, as well as the detection efficiency, of a p-Terphenyl scintillator are lower in comparison with a NE-213 equivalent detector.

Found 

Are you a researcher?

Create a profile to get free access to personal recommendations for colleagues and new articles.
Share
Cite this
GOST | RIS | BibTex | MLA
Found error?