Development of TR-19 Cyclotron Parameter Settings for Fully Automated Production of Radiometals with Applications in Nuclear Medicine
L. S. Crăciun
1
,
T. R. Eșanu
1
,
Radu Anton Leonte
1
,
Hermann Anton Schubert
1
,
Raul Victor Erhan
1
,
Dana Niculae
1
1
Horia Hulubei National Institute for Research and Development in Physics and Nuclear Engineering, Reactorului 30, 077125 Măgurele, Romania
Publication type: Journal Article
Publication date: 2025-02-26
Abstract
At the Radiopharmaceutical Research Center (CCR) of the Horia Hulubei National Institute for R&D in Physics and Nuclear Engineering (IFIN-HH), we operate a TR-19 cyclotron for radio isotope production. To broaden our spectrum of radioisotopes with applications in nuclear medicine, we add a new external beam line towards a state-of-the-art solid target station. Besides practical experience with the implementation of the Comecer ALCEO metal solid targetry system, a new, more efficient way of tuning the beam onto the target and the design of a dedicated neutron local layered shielding are presented.