American Journal of Tropical Medicine and Hygiene, volume 112, issue 3, pages 685-691

Evaluation of the Performance of Portable Hemoglobinometers at Measuring Hemoglobin and Detecting Anemia in a Periurban Pediatric Population in Lima, Peru

Zach Silverstein 1, 2
Alicia Alva Mantari 3, 4
Kimberly Breglio 5
Monica J Pajuelo 3, 6
Gladys Valdiviezo 3
Sandra Cruz 3
Pamela Castañeda 3
Jessica D Rothstein 1, 7
Lucia Huaman-Fernandez 3
Lilia Cabrera 6
Mario Salguedo 3
Franklin Barrientos 3
Bryan Saldivar-Espinoza 3
Dennis Núñez-Fernández 3
Robert H. Gilman 1
Patricia Sheen 3
Avid Roman-Gonzalez 4
Sassan Noazin 1
Mirko Zimic 3
Show full list: 19 authors
Publication typeJournal Article
Publication date2025-03-05
scimago Q1
SJR0.834
CiteScore6.2
Impact factor1.9
ISSN00029637, 14761645
Abstract
ABSTRACT.

As many as one in three people worldwide have anemia, with young children at increased risk of both disease and complications. In settings without clinical laboratories, portable hemoglobinometers serve important roles in diagnosing anemia and estimating prevalence. Here, we assess the validity of two such point-of-care devices—the HemoCue Hb201 and the HemoCue Hb301—relative to the international reference standard, the cyanmethemoglobin method. In total, 428 children ages 6–60 months were recruited at health posts in Lima, Peru, and venous and capillary blood samples were collected from each participant. Venous blood was assessed with the cyanmethemoglobin method, whereas capillary blood was assessed using the Hb201 and the Hb301; 16.1% of participants were found to have anemia using the cyanmethemoglobin method. Both the Hb201 (43.7%) and the Hb301 (20.6%) overestimated this prevalence, with the former reaching statistical significance (P <0.0001 and P = 0.11, respectively). Both devices also tended to underestimate hemoglobin concentration, with the Hb201 (mean difference = −0.99 g/dL; percentage error = −8.1%) being appreciably less accurate than the Hb301 (mean difference = −0.35 g/dL; percentage error = −2.7%). Areas under the curve for the Hb201 (0.92) and the Hb301 (0.93) were statistically similar (P = 0.28); however, the Hb201 incorrectly classified 29.4% of participants compared with 11.0% for the Hb301. Both devices had more false positives than false negatives. In conclusion, the Hb301 was found to be significantly more accurate than the Hb201 at measuring hemoglobin, diagnosing anemia, and estimating anemia prevalence.

  • We do not take into account publications without a DOI.
  • Statistics recalculated only for publications connected to researchers, organizations and labs registered on the platform.
  • Statistics recalculated weekly.

Are you a researcher?

Create a profile to get free access to personal recommendations for colleagues and new articles.
Share
Cite this
GOST | RIS | BibTex | MLA
Found error?