Different Models of Dual-Energy Bone DXA Scanners: A Comparative Study

Publication typeJournal Article
Publication date2022-06-28
scimago Q4
wos Q4
SJR0.147
CiteScore0.8
Impact factor0.1
ISSN23112905, 25420933
Abstract

Background. Dual-energy X-ray absorptiometry (DXA) is an effective method for bone mineral density (BMD) and subcutaneous fat percentage estimation. The constant development of new densitometry techniques, the demographic change and the higher potential of artificial intelligence in healthcare enhance requirements for the high-quality measurements in DXA. This study aimed to develop a quality control method for DXA scanners and compare four DXA systems with different X-ray geometries and manufacturers when simulating fat-water environments. Methods. We evaluated the accuracy (relative error (%) and precision (CV%)) of the bone mineral density (BMD) measurements, performed by the four DXA scanners: 2 with narrow-angle fan beam (64- and 16-channel detectors (DXA-1, DXA-2)); 1 with wide-angle fan beam (DXA-3); 1 with pencil beam (DXA-4). We used a PHK (PHantom Kalium) designed to imitate spine. The PHK contained four vertebras filled with a K2HPO4 solution in various concentrations (50-200 mg/ml). The PHK also included paraffin patches (thickness 40 mm) to simulate the fat layer. Results. DXA-1 and DXA-2 demonstrated the best CV% ranged from 0.56% to 1.05%. The least % was observed when scanning PHK with fat layer on DXA-1 and DXA-2 (1.74% and 0.85%) and DXA-4 (1.47%). DXA-3 produced significantly lower BMD ( = -14.56%, p = 0.000). After removing the fat layer, we observed reduction (p = 0.000) of BMD for DXA- 1 and DXA-2 ( = -5.11% and -6.12% respectively) and weak deviation (p = 0.80) for DXA-4 (0.87%). For DXA-3, removal of the fat layer also resulted in a significant reduction in BMD ( = -16.44%, p = 0.000). The subcutaneous fat modeling showed that all these DXA systems automatically determine the percentage of fat in the scanned area with weak underestimation: for DXA-1, DXA-2 and DXA-4 the % were -5,9%, -6,3% and -2,3% respectively. CV% were 0.15%; 0.39%; 1.6%, respectively. Conclusions. We proved a significant underestimation of the BMD measurements across the entire range of simulated parameters for the DXA scanners when the model did not include the subcutaneous fat layer. All models demonstrated high accuracy in measuring the fat layer, with the exception of the DXA-3 model, which was not assessed in these studies.

Found 
Found 

Top-30

Journals

1
Physics of Atomic Nuclei
1 publication, 100%
1

Publishers

1
Pleiades Publishing
1 publication, 100%
1
  • We do not take into account publications without a DOI.
  • Statistics recalculated weekly.

Are you a researcher?

Create a profile to get free access to personal recommendations for colleagues and new articles.
Metrics
1
Share
Cite this
GOST |
Cite this
GOST Copy
Petraikin A. V. et al. Different Models of Dual-Energy Bone DXA Scanners: A Comparative Study // Traumatology and Orthopedics of Russia. 2022. Vol. 28. No. 2. pp. 48-57.
GOST all authors (up to 50) Copy
Petraikin A. V., Akhmad E., Semenov D. S., Artyukova Z. R., Kudryavtsev N. D., Petraikin F. A., Nisovstova L. A. Different Models of Dual-Energy Bone DXA Scanners: A Comparative Study // Traumatology and Orthopedics of Russia. 2022. Vol. 28. No. 2. pp. 48-57.
RIS |
Cite this
RIS Copy
TY - JOUR
DO - 10.17816/2311-2905-1731
UR - https://journal.rniito.org/jour/article/view/1731
TI - Different Models of Dual-Energy Bone DXA Scanners: A Comparative Study
T2 - Traumatology and Orthopedics of Russia
AU - Petraikin, Alexei V.
AU - Akhmad, Ekaterina
AU - Semenov, Dmitriy S.
AU - Artyukova, Z. R.
AU - Kudryavtsev, N D
AU - Petraikin, F. A.
AU - Nisovstova, L. A.
PY - 2022
DA - 2022/06/28
PB - Eco-Vector LLC
SP - 48-57
IS - 2
VL - 28
SN - 2311-2905
SN - 2542-0933
ER -
BibTex |
Cite this
BibTex (up to 50 authors) Copy
@article{2022_Petraikin,
author = {Alexei V. Petraikin and Ekaterina Akhmad and Dmitriy S. Semenov and Z. R. Artyukova and N D Kudryavtsev and F. A. Petraikin and L. A. Nisovstova},
title = {Different Models of Dual-Energy Bone DXA Scanners: A Comparative Study},
journal = {Traumatology and Orthopedics of Russia},
year = {2022},
volume = {28},
publisher = {Eco-Vector LLC},
month = {jun},
url = {https://journal.rniito.org/jour/article/view/1731},
number = {2},
pages = {48--57},
doi = {10.17816/2311-2905-1731}
}
MLA
Cite this
MLA Copy
Petraikin, Alexei V., et al. “Different Models of Dual-Energy Bone DXA Scanners: A Comparative Study.” Traumatology and Orthopedics of Russia, vol. 28, no. 2, Jun. 2022, pp. 48-57. https://journal.rniito.org/jour/article/view/1731.